• CAPNOSTAT ® 5 CO2 SENSOR:
    • Principle of operation: Main sensor non-dispersive infrared absorption (NDIR) single beam optics logométrico dual wavelength.
    • Response time: Less than 60 ms
    • Effects of gas composition: O2, N2O (operator selectable)
    • Weight: Less than 18 g cordless
    • Sensor size: 1.3 x 1.67 x 0.85 inches. (3.3 x 4.2 x 2.2 cm), cable 8 ft (2.44 m)
    • Construction: Durable plastic high-performance, cable ultra impact resistant: Shock resistance: The sensor supports a fall of 6 feet (1.83 m) on a tile floor. IPX4 rating.
    • Final breath CO2: Limits: 0-150 mmHg, kPa or 0-20% with Pb 760 mmHg
    • Accuracy: ± 2 mmHg to 0-40 mmHg ± 5% of reading to 41-70 mmHg, ± 8% of reading 71-100 mmHg, ± 12% of reading 101-150 mmHg
    • Respiratory: Limits: 2-150 resp. / Min
    • Accuracy: ± 1 resp. / Min

 

  • CO2 SENSOR LoFlo C5:
    • Principle of operation: Main sensor non-dispersive infrared absorption (NDIR) single beam optics logométrico dual wavelength.
    • Response time: Less than 60 ms
    • Effects of gas composition: O2, N2O (operator selectable)
    • The module weighs less than 0.5 lbs (227 grams)
    • Module Size: 3.28 x 2.46 x 1.46 inches [83.3 x 62.5 x 37.1 mm], 20 inches (0.5 m)
    • Construction: Durable plastic high-performance, cable ultra impact resistant. Rating IPX4Final breath CO2: Limits: 0-150 mmHg, kPa or 0-20% with Pb 760 mmHg
    • Accuracy: ± 2 mmHg to 0-40 mmHg ± 5% of reading to 41-70 mmHg, ± 8% of reading 71-100 mmHg, ± 12% of reading 101-150 mmHg
    • Respiratory: Limits: 2-150 resp. / Min
    • Accuracy: ± 1 resp. / Min
 